MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL TRUESPAN MENISCAL REPAIR SYSTEM PEEK 24 DEGREE; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL TRUESPAN MENISCAL REPAIR SYSTEM PEEK 24 DEGREE; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 228152
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2022
Event Type  malfunction  
Event Description
It was reported by a healthcare professional in china that a meniscal repair procedure on (b)(6) 2022, it was observed that the needle on the truespan meniscal repair system peek 24 degree was deformed.Another like device was used to complete the surgery.There were no adverse consequences to the patient nor surgical delay reported.No additional information could be provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary : the complaint device has not been returned, therefore unavailable for a physical evaluation.However, a photo was provided.Upon visual inspection of the photo, the device shaft was observed deformed; therefore, this complaint can be confirmed.A manufacturing record evaluation was performed for the finished device lot number: 9l46759, and no nonconformances were identified.The photo does not provide enough evidence to determine root cause.Without physical evaluation of the device reported, we cannot establish a root cause for the issue experienced by the customer.The possible root cause can be attributed to procedural variables, such handling of the device or product interaction during procedure.As per ifu, during needle insertion use a malleable graft retractor or slotted cannula to prevent the needle from catching on or damaging soft tissue.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received and evaluated; visual inspection reveals that the needle is bent.The plastic sleeve was removed to verify the presence of both plates, the plates were not found inside the applier needle and they were not returned.A manufacturing record evaluation was performed for the finished device 9l46759 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection and the functional test results, this complaint can be confirmed.A possible root cause can be attributed to procedural variables, such handling of the device or product interaction during procedure; an excessive manipulation of the device, tilting movements of the device while it was inserted causing the needle to be deformed, however this cannot be conclusively determined.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUESPAN MENISCAL REPAIR SYSTEM PEEK 24 DEGREE
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15836140
MDR Text Key307597454
Report Number1221934-2022-03658
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705026029
UDI-Public10886705026029
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K153667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number228152
Device Catalogue Number228152
Device Lot Number9L46759
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-