MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VASOVIEW HEMOPRO 2
Device Problems Mechanical Problem (1384); Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/28/2022
Event Type  malfunction  
Event Description
The hospital reported that during an endoscopic vein harvesting procedure using vasoview hemopro 2.About halfway through the case, the tool button was sticking and not activating the cut and seal feature.They had to open up another kit to finish the case.There were no patient affects.
 
Manufacturer Narrative
Trackwise id (b)(6).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
N/a.
 
Manufacturer Narrative
(b)(4).The device was returned to the factory for evaluation on 11/11/2022.An investigation was conducted on 11/17/2022.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the heater wire or the clear silicone insulation on both the cold and hot jaws.A mechanical evaluation was conducted.The blue toggle was manipulated to open and close the jaws.A little resistance was observed on the toggle while manipulation of the toggle, however the jaws were able to opened and closed.An electrical evaluation was conducted.A pre-cautery test was performed per the direction for use (dfu) with a reference cable, adapter and reference power supply vh-3010 at level 3.0.The device passed the pre-cautery test; it produced visible steam and heat during ten (10) 3-second activations and shut off when the toggle was released.To evaluate the safety shut down system, a polyfuse activation test was performed 5 times over 10 minutes.The device shut off after the period of sustained activation and reactivated after 10-second cooling period with no incident each time.An activation and transection capability test was performed over four (4) repetitions using "max life test method stm2048073 rev aa.The device successfully transected tissue four (4) times.A temperature and resistance test was conducted to evaluate the device function per hemopro 2 final test 90523436 rev w.The resistance value was measured at.68 ohms which is within specification.The device passed the temperature measurements test.The displayed temperature increased and turned green within the 2 second specified timeframe.The displayed temperature decreased once the toggle swivel was released.Based on the results of the evaluation, the complaint for the reported failure "failure to cut¿ was not confirmed, but the reported failure "mechanical problem" was observed.The lot #25161422 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key15838692
MDR Text Key306518828
Report Number2242352-2022-00953
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/04/2023
Device Model NumberVASOVIEW HEMOPRO 2
Device Catalogue NumberC-VH-4000
Device Lot Number25161422
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/19/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK.; UNKNOWN.
-
-