Catalog Number 2C4041 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/07/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that mold was observed in eight (8) y-type tur/bladder irrigation sets.The issue was identified before patient use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Device manufacturer address 1: 600 mts.(b)(4) parque industrial.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: eight (8) samples were received for evaluation.A visual inspection was performed with the naked eye to all of the samples, which observed particles outside the fluid path between the walls of the tubing in two (2) of the samples.These particles were noted to be intrinsic to the raw material of the device.The reported condition was verified in two of the samples only.The cause of the condition was determined to be manufacturing related.No issues were observed in the other six (6) samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|