• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 367955
Device Problem Missing Information (4053)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/04/2022
Event Type  malfunction  
Event Description
It was reported when using the bd vacutainer® sst¿ ii advance plus blood collection tubes had no label.The following information was provided by the initial reporter.The customer stated: ¿no label.There is a label on the main packaging.There are no labels on the tubes.That's why the institution destroyed the test tubes.Such a problem was detected in only 1 package of products.".
 
Manufacturer Narrative
Investigation summary: bd did receive 2 photographs from the customer in support of this complaint.An evaluation of the attached photographs shows evidence of missing label on the tube.Additionally, 100 retained samples from the bd inventory were visually checked and no missing labels were found.Bd was able to confirm the customer¿s indicated failure mode based on the photographs provided, however retained samples were satisfactory.Bd was not able to identify a root cause for the indicated failure mode.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15848334
MDR Text Key307709378
Report Number9617032-2022-01179
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903679558
UDI-Public50382903679558
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
BK050036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date11/30/2023
Device Model Number367955
Device Catalogue Number367955
Device Lot Number2145912
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-