• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Insufficient Flow or Under Infusion (2182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/24/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customers allegation was confirmed.It was found that the dampener in the device was damaged.The suction tube was also deformed.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
It was reported to olympus that the high flow insufflation unit had internal loosness.Upon inspection and testing of the returned loaner device, it was noted that the maximum gas flow wasn't up to the standard due to a faulty first regulator unit.There were no reports of patient harm associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see the updates in sections d4, d8, e4, h4, h6, and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely due to a failure of the primary regulator unit.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyamaodakura
nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15848756
MDR Text Key307698479
Report Number3002808148-2022-04550
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-