This report has been identified as b.Braun medical internal report number (b)(4).The affected device was provided for further evaluation.Visual inspection confirmed two #8 labels were found, one of them in place of #9 which was found to be missing.The reported defect as confirmed.Based on the evaluation performed, the defect derives from an accidental human mistake happening during the manual assembly.As a result of this occurrence, retraining was performed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|