Model Number 27977 |
Device Problem
Inappropriate or Unexpected Reset (2959)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/1901 |
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
|
Event Description
|
The customer reported the device turns off and on randomly.No patient impact or consequences were reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: the returned device was evaluated.The device intermittently emitted a tone and powered off on its own as if the power button was pressed.Internal inspection of the device determined contamination on the circuit board power button resulted in the device powering on and off by itself.
|
|
Event Description
|
The customer reported the device turns off and on randomly.No patient impact or consequences were reported.
|
|