Model Number 2020-00-401 |
Device Problem
Device-Device Incompatibility (2919)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/01/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional procode: gea.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from canada reports an event as follows: it was reported that on (b)(6) 2022, the symphony screwdrivers in question were binding up during assembly.On closer inspection it was found to be difficult to pass the driver shaft through the driver sleeve.There were no reported patient consequences.No further information is available.This report is for a symphony oct system polyaxial screw driver retention sleeve.This is report 6 of 8 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A review of the receiving inspection (ri) for poly screw driver reten sleeve was conducted identifying that lot number (b)(4) was released in one batch.Batch1: lot qty of (b)(4) units were released on 03 mar 2020 with no discrepancies.Supplier: depuy : greatbatch medical.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A review of the receiving inspection (ri) for poly screw driver reten sleeve was conducted identifying that lot number (b)(4) was released in one batch.Batch1: lot units were released on 03 mar 2020 with no discrepancies.Supplier: depuy : (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the poly screw driver reten sleeve only presents signs of normal wear.A dimensional inspection for the poly screw driver reten sleeve was not performed since it is not applicable to the complaint condition.A functional test was performed, the retention sleeve was assembled with the poly screw driver shft x20 lot number , the shaft was able to be secured using the button locking mechanism with no signs of issue.The complaint condition was not able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the poly screw driver reten sleeve was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: the following drawings reflecting the current and manufactured revisions were reviewed: - 103140293 poly screwdriver retention sleeve assy rev.E / rev.C.Dimensional inspection: n/a.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|