• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754WWS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754WWS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-754WWS
Device Problems Mechanical Problem (1384); Environmental Compatibility Problem (2929)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/15/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information received by medtronic indicated that the insulin pump was exposed to high magnetic environment.Customer also reported motor error alarm and was able to clear the alarm and able to rewind.No harm requiring medical intervention was reported.The device will be returned for analysis.
 
Manufacturer Narrative
Pump alarmed motor error during rewind.Unable to perform the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test and excessive no delivery test due to motor error alarms.Pump history download using thds was successful.However, a47 alarms found on the history file which caused corrupted data.Unable to confirm the reported event date due to the corrupted data.A47 alarms are due to the pump not having power for a long period of time.The following were noted during visual inspection: cracked belt clip slot, stained end cap sticker, stained address/serial number label and cracked reservoir tube lip.The test p-cap/reservoir does lock into place.Unit received with motor error alarm during rewind due to motor encoder signal out of phase.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754WWS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key15879190
MDR Text Key307302485
Report Number2032227-2022-366156
Device Sequence Number1
Product Code OYC
UDI-Device Identifier000000763000161286
UDI-Public(01)000000763000161286
Combination Product (y/n)N
Reporter Country CodeTU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-754WWS
Device Catalogue NumberMMT-754WWS
Device Lot NumberB8754WWSJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received07/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-