Catalog Number ECH060020W
Device Problem Peeled/Delaminated (1454)
Patient Problem Stenosis (2263)
Event Date 10/25/2022
Event Type  Injury  
Manufacturer Narrative
Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
The following was reported to gore: on (b)(6) 2022, a patient was implanted with an acuseal vascular graft (vg) to treat arteriovenous fistula from brachial artery to axillary vein.The patient was treated hemodialysis through acuseal vg in the evening on the same day.On (b)(6) 2022, physician found that there was thrombus through the acuseal vg at puncture site of hemodialysis treatment, and he suspected that there was vg delamination.On (b)(6) 2022, physician found that there was more organized thrombus at puncture site of hemodialysis treatment.On (b)(6) 2022, patient was diagnosed with vg occlusion, therefore, physician performed surgery and removed acuseal vg.Another non-gore vascular graft was used to replace the removed one successfully.Patient tolerated the procedure.
 
Manufacturer Narrative
H3 - device evaluated by manufacturer - yes a review of the manufacturing records indicated the lots met all pre-release specifications.Product evaluation: the device fragments were returned to w.L.Gore & associates for investigation.Submitted in formalin was one gore-tex stretch vascular graft fragment attached to one gore acuseal vascular graft fragment (vgf-1) via graft-to-graft anastomosis and one gore acuseal vascular graft fragment (vgf-2).The device fragments had been transected prior to arrival at w.L.Gore & associates.The lumens of both fragments were widely patent.On vgf-1, the lumen was patent but significantly narrowed through the fragment length by brown/tan soft tissue.On vgf-2, a very small amount of gelatinous brown material was present within the lumen.Also, on vgf-2 near extremity b, there was a partially circumferential, thin, translucent membrane of inner eptfe reflected into the lumen.The abluminal surfaces were generally devoid of tissue.Histopathologic analysis was not performed, due to the lack of tissue on the fragments.The fragments were subjected to an enzymatic digestion process to remove biologic debris.Following digestion all fragments were examined for material disruptions with the aid of a stereomicroscope.Outer eptfe and silicone layer separation was present on both fragments and was consistent with separation caused by interaction with cutting and grasping/pulling with surgical instrumentation (i.E., forceps/clamps, scissors, scalpel), likely used during a surgical procedure.On vgf-1, inner eptfe and silicone layer separation was present in areas, which were likely caused by shallow-angle cannulation events.Cannulation marks were present within 2 mm of the graft-to-graft anastomosis on vgf-1.Inner eptfe and silicone layer separation was also present on vgf-2.Due to the limited/short length of the fragment returned and presence of other surgically induced material disruptions (e.G., transections, forceps marks) on vgf-2, the cause of the inner eptfe and silicone layer separation could not be determined with the information provided.
 
Event Description
The following was reported to gore: on (b)(6 2022, a patient was implanted with an acuseal vascular graft (vg) to treat arteriovenous fistula revision from brachial artery to axillary vein.The patient had been implanted with a gore-tex® stretch vascular graft in months ago.Physician performed a jumping graft bypass which was bypass another graft.One end to end anastomosis to arterial side and anther one doing a anastomosis to venous.The patient was treated hemodialysis through acuseal vg in the evening on the same day.On (b)(6) 2022, physician found that there was thrombus through the acuseal vg at puncture site of hemodialysis treatment, and he suspected that there was vg delamination.On (b)(6) 2022, physician found that there was more organized thrombus at puncture site of hemodialysis treatment.On (b)(6) 2022, patient was diagnosed with vg occlusion, therefore, physician performed surgery and removed acuseal vg.A part of gore-tex® stretch vascular graft was removed together with the acuseal vg via a graft-to-graft anastomosis.There was no product information nor complaint for gore-tex® stretch vascular graft.Another non-gore vascular graft was used to replace the removed one successfully.Patient tolerated the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® ACUSEAL VASCULAR GRAFT
Type of Device
PROSTHESIS, VASCULAR GRAFT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WEST B/P
1505 n. fourth street
flagstaff AZ 86004
Manufacturer Contact
qiong wang
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15891394
MDR Text Key304587990
Report Number2017233-2022-03521
Device Sequence Number1
Product Code DSY
Combination Product (y/n)Y
Reporter Country CodeTW
PMA/PMN Number
K130215
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberECH060020W
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient SexMale
Patient Weight80 KG
-
-