MEDTRONIC NEUROMODULATION EXTERNAL NEUROSTIMULATOR; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number NEU_ENS_STIMULATOR |
Device Problem
Malposition of Device (2616)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
Injury
|
Manufacturer Narrative
|
Continuation of concomitant medical products: product id neu_unknown_lead lot# unknown serial#, implanted: explanted: product type lead product id neu_unknown_lead, lot# unknown, serial# , implanted: explanted: product type lead section d information references the main component of the system.Other relevant device(s) are: product id: neu_unknown_lead, serial/lot #: unknown, ubd: , udi#: ; product id: neu_unknown_lead, serial/lot #: unknown, ubd: , udi#: medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that both the patient's sensight electrodes were inserted off target due to equipment error and they needed to prepare for a revision.The physician would use their own tools instead of the insertion tool.
|
|
Manufacturer Narrative
|
H10: review of this mdr and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received: it was reported that the issue of the issue was caused by an error with the rosa robot.There were no issues with sensight or medtronic involved.***this event is no longera reportable event.Mdr decision corrected to not reportable.No additional supplemental mdrs are required unless additional information received makes the event reportable.Due to gch functionality, the complaint flag cannot be flipped to 'no' as a regulatory report exists in this pe***.
|
|