Catalog Number UNKNOWN G2 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Chest Pain (1776)
|
Event Date 11/09/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
As the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.
|
|
Event Description
|
It was reported that fifteen years post filter placement, the device allegedly fractured with fragments retained in right ventricle and right pulmonary artery.It was further reported that the filter was removed with forceps and retained fragments removed with forceps and snares.Reportedly, patient experienced chest pain.The current status of the patient is unknown.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the alleged detachment of device or device component could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: g3 h11: b5, g2 h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that approximately fifteen years and two months post filter placement, the device allegedly fractured with fragments retained in right ventricle and right pulmonary artery.It was further reported that the filter was removed with forceps and retained fragments removed with forceps and snares.Reportedly, patient experienced chest pain.The current status of the patient is unknown.
|
|