C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problem
Arrhythmia (1721)
|
Event Date 11/18/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refz3340 showed four other similar product complaint(s) from this lot number.
|
|
Event Description
|
It was reported by the customer "has informed me that they have had 3 instances over the past few weeks of 3cg confirming good picc location, and then upon x-ray it appears a few centimeters off from the lower svc.The 3 patients have had ectopies and runs of v tach within a day after picc placement, when the 3cg readings/p-waves looked really good during placement." this report addresses the second device.
|
|