• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Arrhythmia (1721)
Event Date 11/18/2022
Event Type  Injury  
Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refz3340 showed four other similar product complaint(s) from this lot number.
 
Event Description
It was reported by the customer "has informed me that they have had 3 instances over the past few weeks of 3cg confirming good picc location, and then upon x-ray it appears a few centimeters off from the lower svc.The 3 patients have had ectopies and runs of v tach within a day after picc placement, when the 3cg readings/p-waves looked really good during placement." this report addresses the second device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15929188
MDR Text Key304955785
Report Number3006260740-2022-05587
Device Sequence Number1
Product Code LJS
UDI-Device Identifier00801741213304
UDI-Public(01)00801741213304
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K180548
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Model NumberN/A
Device Catalogue NumberCK000992
Device Lot NumberREFZ3340
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-