• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ENDURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ENDURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2172
Device Problem Over-Sensing (1438)
Patient Problem Dizziness (2194)
Event Date 11/21/2022
Event Type  Injury  
Event Description
Related manufacture reference number: 2017865-2022-48273; related manufacture reference number: 2017865-2022-48274.It was reported that the patient presented during clinical follow-up.Upon interrogation, it was noted pacemaker and right ventricular exhibited noise oversensing causing pacing inhibition and dizziness.It was also noted that the atrial lead exhibited high capture threshold.The pacing system was explanted and replaced on (b)(6) 2022.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of noise over-sensing was not confirmed.As received, a device was returned above elective replacement indicator (eri) level for analysis.Electrical test and longevity assessment did not identify any anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15951026
MDR Text Key305174496
Report Number2017865-2022-48272
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734509602
UDI-Public05414734509602
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model NumberPM2172
Device Catalogue NumberPM2172
Device Lot NumberA000085448
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-