(b)(4).Concomitant medical products: femoral hinge service kit xt size c, use with distal femoral xt size c, use with xt only.Catalog # 00585007113, lot # 65329432.Device evaluated by mfr: the complaint device has been returned, but the device investigation has not yet been completed.Once the evaluation is completed, a supplemental medwatch will be submitted.
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g4, g7, h1, h2, h3, h6, h9, h10.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows signs of implantation nicked, gouged, and worn.Additionally, the locking features were damaged/compressed.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|