• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number NX1000-3
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Insufficient Information (4580)
Event Date 11/06/2022
Event Type  Death  
Manufacturer Narrative
The device was received for evaluation and successfully passed testing.All devices must meet all quality criteria and manufacturing specifications prior to release.A review of the device history record (dhr) was conducted which confirmed that the device met all quality criteria and manufacturing specifications prior to release.Available log files were retrieved and analyzed which showed device performance was without deficiency and was unremarkable.There was no indication of a device malfunction from the available information.The nxstage system one user guide supplement for solo home hemodialysis outlines risks, warnings, and requirements for the solo home hemodialysis patient.Udi: (b)(4).
 
Event Description
A report was received on 16 nov 2022 from the home therapy nurse (htn) of a 42 year old male patient with a medical history of end stage renal disease and who was approved for performing solo home hemodialysis therapy, stating the patient expired during a hemodialysis session on (b)(6) 2022.Additional information was received on 21 nov 2022 from the htn confirmed the patient was treating without a caregiver present when he was found expired with his venous needle dislodged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key15954022
MDR Text Key305234926
Report Number3003464075-2022-00083
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberNX1000-3
Device Catalogue NumberCHRONIC HI-FLOW CYCLER
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2022
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/26/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age42 YR
Patient SexMale
-
-