Model Number RT266
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint rt266 infant dual heated evaqua2 breathing circuit is currently en route to fisher & paykel healthcare (f&p) in new zealand for evaluation.We will provide a follow up report upon completion of our investigation.
 
Event Description
A healthcare facility in california reported via a fisher & paykel healthcare (f&p) field representative that a rt266 infant dual heated evaqua2 breathing circuit was leaking during patient use from the connection of the dryline and the inspiratory filter.There was no patient consequence.
 
Manufacturer Narrative
(b)(4).Method: the complaint rt266 infant dual-heated evaqua2 breathing circuit was returned to fisher & paykel healthcare (f&p) in new zealand, where it was visually inspected and leak tested.Results: visual inspection of the complaint rt266 infant dual-heated evaqua2 breathing circuit revealed no damage was found to the dryline.The dryline port cap was observed not fully inserted into the port.The leakage test also revealed that the breathing circuit was within specification.Conclusion: we were unable to determine what may have caused the failed leak test as reported by the customer, as no fault was found with the returned device.However, the leak the customer experienced was most likely due to the dryline port not fully closed.All rt266 infant dual-heated evaqua2 breathing circuits are visually inspected, and pressure and flow tested during production, and those that fail are rejected.The subject infant breathing circuit would have met the required specifications at the time of production.Our user instructions that accompany the rt266 infant dual-heated evaqua2 breathing circuit state the following: - "check all connections are tight before use." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
 
Event Description
A healthcare facility in california reported via a fisher & paykel healthcare (f&p) field representative that a rt266 infant dual heated evaqua2 breathing circuit was leaking during patient use from the connection of the dryline and the inspiratory filter.There was no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key15966102
MDR Text Key305510435
Report Number9611451-2022-01175
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT266
Device Catalogue NumberRT266
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received04/18/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-