• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fatigue (1849); Muscle Weakness (1967); Pain (1994); Dizziness (2194); Ambulation Difficulties (2544); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 12/10/2022
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 8709sc, lot# n244287009, implanted: (b)(6) 2010, product type catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from multiple sources, company representative (rep), consumer (con), healthcare provider (hcp) regarding a patient receiving lioresal 750 mcg/ml at 234.7 mcg/day via an implantable pump.It was reported the patient's pump was in early replacement indicator (eri) and had been hospitalized.The patient reported having increased pain, dizziness, fatigue, weakness, and difficulty ambulating.It was reported surgical intervention was planned but was yet to be scheduled.Patient weight and medical history were asked but remain unknown.At the time of this report the issue had not been resolved and the patient's status was alive - no injury.
 
Event Description
Additional information received from the healthcare provider via a company representative reported that there was no device malfunction, the pump had reached eri.The pump has not yet been replaced.
 
Manufacturer Narrative
Concomitant medical products: product id 8709sc lot# n244287009 implanted: (b)(6) 2010 product type catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15978568
MDR Text Key305489274
Report Number3004209178-2022-16409
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169508156
UDI-Public00643169508156
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/28/2017
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Date Manufacturer Received12/14/2022
Date Device Manufactured02/02/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H10."
Patient Outcome(s) Hospitalization;
Patient Age59 YR
Patient SexMale
-
-