• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBE; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBE; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 367955
Device Problem Premature Separation (4045)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2022
Event Type  malfunction  
Event Description
It was reported when using the bd vacutainer® sst¿ ii advance plus blood collection tube there was stopper pull out of the tube when in an analyzer.This event occurred 2 times.The following information was provided by the initial reporter.The customer stated: "tubes are opening.Customer had two events with yellow cap tubes in centrifuge.".
 
Manufacturer Narrative
Investigation summary: bd had not received samples or photos for investigation.Therefore, eight retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to stopper pop off as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode stopper pop off.Bd was not able to identify a root cause for the indicated failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBE
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15982642
MDR Text Key307690426
Report Number9617032-2022-01260
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903679558
UDI-Public50382903679558
Combination Product (y/n)N
Reporter Country CodeCI
PMA/PMN Number
BK050036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/31/2023
Device Model Number367955
Device Catalogue Number367955
Device Lot Number2056918
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-