Oem manufacturer: the manufacturing location for this product is nipro.This site is an oem manufacturing site.(b)(4).Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: bd had not received samples or photos for evaluation.Additionally, bd was unable to determine the specific lot number associated with this complaint; therefore, a review of the device history record could not be conducted.This complaint is unable to be confirmed.If additional information is made available, this complaint will be reopened to assess the level of investigation needed.During our manufacturing process, molding parts such as the safety shields or the wing needle hubs are inspected per a lot and only the conforming parts will be used for the assembly.In addition, the appearance and the functionality are checked for all product lots as process inspections and release inspections.Therefore, the possibility that defective products are flowed to the next process or shipped out is very low.As stated in the ifu, please hold the end of safety shield and tubing (avoid holding the wings together) and slide the safety shield straight to cover the winged needle without holding wings.Under our manufacturing process, inspections are performed for all the products by automatic blockage inspection machine installed on our automatic assembling machine after adhering luer adaptor hubs and needle tubes, and after adhering wing needle hubs, tubes and connecters.Moreover, inspection machines are routinely checked and maintained.Therefore, the possibility that clogged products are flowed and shipped out is very low.
|