• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, CROSSFIRE 2 CONSOLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, CROSSFIRE 2 CONSOLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0475100000
Device Problem Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2022
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that the device sparked.
 
Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.Alleged failure: did not pass bio med all sparked whrn plugged in the root cause is the rf board.The failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that the device sparked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, CROSSFIRE 2 CONSOLE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15987385
MDR Text Key306994661
Report Number0002936485-2022-00743
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071859
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0475100000
Device Catalogue Number0475100000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received11/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-