• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM (ACTIVE SENTRY); UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM (ACTIVE SENTRY); UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065753057
Device Problems Power Problem (3010); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A customer reported that an ophthalmic operating console presented problems with alternating current (ac) unplugging from wall and console shuts down.The procedure details and patient impact were not provided.Additional information has been requested but none received till date.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The company representative confirmed and replicated the reported event.The nonconforming multifunctioning in/out (mfio) printed circuit board (pcb) was replaced to resolve the issue.The system was tested and found to meet product specifications.A non-conformance-based review of the batch/lot/serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.The root cause of the reported event is attributed to the nonconforming mfio pcb.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM (ACTIVE SENTRY)
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15996283
MDR Text Key308073334
Report Number2028159-2022-01770
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065753057
Device Lot Number13N94R
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-