• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2272
Device Problems Break (1069); Loose or Intermittent Connection (1371); Misassembled (1398)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2022
Event Type  malfunction  
Event Description
Related manufacturer reference number: 2017865-2022-48923.It was reported that the patient presented for an implant procedure.During the procedure, the right ventricle (rv) lead was incorrectly connected to the atrial port.The pacemaker set screw was stripped when it was tightening.Multiple attempts and multiple different hex wrenches failed to fit the pacemaker set screw.The rv lead was successfully removed by pulling with a needle driver.The pacemaker and rv lead were damaged and were not used.The physician continued to use the second pacemaker and rv lead to complete the procedure.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16002034
MDR Text Key307094717
Report Number2017865-2022-48922
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPM2272
Device Catalogue NumberPM2272
Device Lot NumberA000132502
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-