Catalog Number 100/800/060CZ
Device Problem Moisture or Humidity Problem (2986)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2022
Event Type  malfunction  
Manufacturer Narrative
No information has been provided to date.A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that there is lots of moisture inside the tube.No injury reported.It was discovered prior to patient use.The outcome of the event was resolved.
 
Manufacturer Narrative
Manufacturing device history record review was not performed because no fault found.Ten (10) unopened product samples were received for evaluation.Visual and functional testing was performed.Visual inspection found the samples appeared to be in good condition.There was not any moisture inside unit packs.Functional testing found the sterile barrier (unit pack) was evaluated as acceptable and there was not found any packaging related nonconformity.Device passed all functional testing.No problem found.Root cause cannot be determined as the sample was successfully tested and no discrepancies were detected.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional investigative action will be taken accordingly.This remediation mdr was generated under protocol (b)(4) as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES BLUE LINE ULTRA (BLU)
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ  753 01
Manufacturer (Section G)
NULL
olomoucka 306
hranice 1, mesto 753 0 1
EZ   753 01
Manufacturer Contact
jim vegel
6000 nathan lane north
seongbuk-gu
minneapolis, MN 55442
MDR Report Key16017020
MDR Text Key306257838
Report Number3012307300-2022-28003
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number100/800/060CZ
Device Lot Number4121503
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2022
Was the Report Sent to FDA? No
Date Manufacturer Received04/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/29/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No