Model Number IPN915191 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/27/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Qn#(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
|
Event Description
|
It was reported that the applier was sent to our technical specialist as the user felt that the clip did not easily leave the applier during a laparoscopic cholecystectomy.No clip fell in the patient.On follow-up, the specialist confirmed that the applier could not be assembled properly; the inner shaft got detached after assembly.
|
|
Event Description
|
It was reported that the applier was sent to our technical specialist as the user felt that the clip did not easily leave the applier during a laparoscopic cholecystectomy.No clip fell in the patient.On follow-up, the specialist confirmed that the applier could not be assembled properly; the inner shaft got detached after assembly.
|
|
Manufacturer Narrative
|
(b)(4).During visual and functional examination, we found that even after many repeats of clip loading, we could not find any failure.The clips were loaded well, closed properly, and could be delivered easily.Upon review of the device history records for the affected lot number, we can confirm that both the correct material and correct components had been used and that the instrument meets the product specifications.All process steps were found to have been properly documented.100% functional test at final inspection found good.We could not find this failure on the instrument; we cannot reproduce the failure.No further measures will be initiated by k & w this complaint is being rejected by us.
|
|
Search Alerts/Recalls
|