Model Number CSE-P-225-JA
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2022
Event Type  malfunction  
Event Description
On (b)(6) 2022, haemonetics was notified of a inner core leak post procedure in japan, utilizing the cell saver® elite® autotransfusion system and cell saver® elite set - 225ml.There was no reported impact to patient health.
 
Manufacturer Narrative
On (b)(6) 2022 the complaint investigation of the returned bowl from the cell saver® elite set - 225ml confirmed a crack in the inner core base.Although there was no injury or harm to the patient, past reporting (b)(4) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of (b)(4) indicated the inner core of the bowl malfunctioned, via a crack, but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due to the event of the past report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER® ELITE® PLUS AUTOTRANSFUSION SYSTEM
Type of Device
CELL SAVER ELITE SET - 225ML
Manufacturer (Section D)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer Contact
brenda bruyere
125 summer street
boston, MA 02110
MDR Report Key16023548
MDR Text Key308120302
Report Number1219343-2022-00060
Device Sequence Number1
Product Code CAC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120586
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCSE-P-225-JA
Device Lot Number0422077
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2022
Date Manufacturer Received11/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-