ST. JUDE MEDICAL, INC. WORKMATE¿ CLARIS¿ DISPLAY PLUS AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE
|
Back to Search Results |
|
Model Number H700150 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/16/2022 |
Event Type
malfunction
|
Event Description
|
From the beginning of the procedure, an error message was noted stating: "amplifier disconnected" on claris and the procedure was cancelled.All connections were checked and the system was restarted (claris and amplifier) several times which did not resolve the issue.Nothing changed, and all devices had power, including the media converters.Two replacement fiber optic cables were tested, but the issue persisted.When the claris was restarted, an error message was noted stating: "could not reconnect all network drives" for approximately two seconds.After an hour and a half, connection could not be established and the procedure was cancelled with no consequences to the patient.
|
|
Manufacturer Narrative
|
Additional information: g9, h2, h3.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported communication issue and subsequent cancellation remain unknown.
|
|