Model Number H700150
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/16/2022
Event Type  malfunction  
Event Description
From the beginning of the procedure, an error message was noted stating: "amplifier disconnected" on claris and the procedure was cancelled.All connections were checked and the system was restarted (claris and amplifier) several times which did not resolve the issue.Nothing changed, and all devices had power, including the media converters.Two replacement fiber optic cables were tested, but the issue persisted.When the claris was restarted, an error message was noted stating: "could not reconnect all network drives" for approximately two seconds.After an hour and a half, connection could not be established and the procedure was cancelled with no consequences to the patient.
 
Manufacturer Narrative
Additional information: g9, h2, h3.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported communication issue and subsequent cancellation remain unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WORKMATE¿ CLARIS¿ DISPLAY PLUS AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16025964
MDR Text Key306330529
Report Number2184149-2022-00305
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K151911
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH700150
Device Catalogue NumberH700150
Device Lot Number6958894
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/16/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-