It was reported when using the bd vacutainer® sst¿ blood collection tubes customer found tubes that had no labels.The following information was provided by the initial reporter.The customer stated: "the doctor ordered dabiochemical, two people checked the doctor's order, checked the barcode, and checked the test tube.When they found that the test tube had no label, they reported it to the head nurse, stopped using it and replaced the test tube, and no adverse effects were caused to the patient.".
|
Investigation summary: bd had not received samples or photos for investigation.Therefore, thirty (30) retain samples were subject to a visual inspection for missing labels on the product.All samples passed testing with no evidence of missing labels on any of the 30 tubes.The device history record was reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.Based on a review of batch records and the investigation results, no root cause from the manufacturing process has been identified as a contributor to the missing label.
|