Model Number CNA0T0
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 11/09/2022
Event Type  malfunction  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.There have been one other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that during a cataract extraction with intraocular lens (iol) implant procedure, a lens loading issue was noted.There was patient contact, and the procedure was completed on the same day.Additional information has been requested.
 
Manufacturer Narrative
The product was returned for analysis and the reported complaint was observed.The device was returned loose in the product carton.The lock-out assembly has been removed.No viscoelastic is observed in the device.The plunger has been advanced at the wound guard and is advanced under the lens.The lens is advanced into the nozzle entry and is crushed.A plunger underride was observed.The root cause may be related to failure to follow the instruction for use (ifu).Inadequate viscoelastic was observed in the device.The ifu instructs: fill the device until ovd can be observed flowing to the nozzle tip.This will require approximately 0.28 ml of viscoelastic.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become stuck in the device allowing the plunger to underride the lens.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16046419
MDR Text Key307777238
Report Number9612169-2022-00718
Device Sequence Number1
Product Code HQL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P190018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCNA0T0
Device Lot Number25415622
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-