• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACUITY X4 SPIRAL S; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACUITY X4 SPIRAL S; IMPLANTABLE LEAD Back to Search Results
Model Number 4675
Device Problem High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2022
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the system with this left ventricular (lv) lead and a cardiac resynchronization therapy pacemaker (crt-p), showed a high, out of range (oor) pacing impedances alert short after being implanted.The impedance value was high and went oor at implant procedure.The lv lead and the crt-p remain in service.It was recommended to reprogram the oor limit for the device.No procedure was recommended.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUITY X4 SPIRAL S
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16046668
MDR Text Key307990003
Report Number2124215-2022-54903
Device Sequence Number1
Product Code LWP
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S398
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number4675
Device Catalogue Number4675
Device Lot Number824127
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
-
-