MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1711K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1711K |
Device Problem
Visual Prompts will not Clear (2281)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/19/2022 |
Event Type
malfunction
|
Event Description
|
Information received by medtronic indicated the customer received frozen screen.No harm requiring medical intervention was reported.Troubleshooting was performed.Product will be returned for analysis.The customer will discontinue to use of the device.
|
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the ngp 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the ngp insulin pump, which is marketed in the united states.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Unit passed the self test.Unable to perform the displacement test due to unresponsive keypad.Unresponsive keypad due to flattened dome switch.(no crease).No unexpected frozen screen or blank display anomaly noted.The history was download successfully using thus software.Unit received with fading serial number label, cracked battery tube threads, and cracked case corner of belt clip rails.The unit p-cap / test reservoir locks in place properly.Unit was not confirmed for frozen display or blank anomalies noted.However, received with unresponsive keypad due to flattened dome switch.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|