Only the lens was returned wrapped in gauze inside the carton.Viscoelastic was dried on the lens.The optic edge was torn and split toward the optic center.One haptic was broken in the gusset area.A qualified viscoelastic was indicated.Broken haptic and torn, split optic damage was observed.A qualified viscoelastic was indicated.It is unknown if the lens and plunger were proper positions for advancement while in the device.Only the lens was returned.The instruction for use (ifu) instructs: after the lens has been advanced to the nozzle line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.After confirming the lens is properly positioned and the haptics are folded properly, proceed with lens implantation.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a broken haptic or other negative outcome, since the haptic may be trapped and stretched, and/or pinched and sheared by the moving plunger.The manufacturer internal reference number is: (b)(4).
|