Model Number FGS-0503
Device Problem Entrapment of Device (1212)
Patient Problems Abdominal Pain (1685); Ulcer (2274)
Event Date 10/17/2022
Event Type  Injury  
Event Description
According to the reporter, the study did not look right.The capsule stayed in patient stomach.Patient contacted the medical doctor's office with complaints of abdominal pain.Doctor brought patient in for an upper endoscopy and found the capsule was lodged in the pylorus from an area of stenosis and the capsule never passed out of this area.The capsule was able to retrieve using a retrieval net and the patient had a small ulcer from the capsule being lodged.The patient was fine and stable after the procedure.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The involved device was not returned.No information regarding the specific customer issue that occurred with the device was available.It was reported that the capsule remained in the patient longer than expected.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: risks associated with capsule ingestion and transit is minimal.The primary hazard is capsule retention.Retention incidence, as determined by a review of published studies of capsule endoscopy in adults, is estimated as 0.75% in patients without known stenosis and 21% in patients with known stenosis.Stenosis and strictures can be complications in inflammatory bowel disease.(doc-4092-01).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTPILL
Type of Device
GASTROINTESTINAL MOTILITY SYSTEM, CAPSULE
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS  20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS   20692
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16054100
MDR Text Key306239345
Report Number9710107-2022-00263
Device Sequence Number1
Product Code NYV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092342
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/22/2023
Device Model NumberFGS-0503
Device Catalogue NumberFGS-0503
Device Lot Number57327P
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-