MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102096-81A
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2022
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event Description
On 28 november 2022, senseonics was made aware of an incident where user received an early sensor replacement alert resulting in an early sensor removal.
 
Manufacturer Narrative
The sensor was investigated.Review of in-vivo data in dms shows that ec#30 (led disconnect error) was triggered on 22 november 2022.In-vitro qc test did not find any performance issue with the sensor.However, sensor manager software measurements test of the sensor showed that the led flag for sensor (b)(6) cleared at field current threshold higher than 412 adc counts which is the root cause of this issue.As part of resolution, the rma was authorized for a sensor replacement.D9 device available for evaluation? yes, device received on 18 jan 2023.H3.Device evaluated by manufacturer?yes.H6.Type of investigation updated to 10.H6.Investigation findings updated to 4203.H6.Investigation conclusions updated to 4307.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key16054666
MDR Text Key307082080
Report Number3009862700-2022-00193
Device Sequence Number1
Product Code QHJ
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/30/2022
Device Model Number102096-81A
Device Catalogue NumberFG-7200-00-301
Device Lot NumberWP08799
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received11/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR