Model Number E161 |
Device Problems
High impedance (1291); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the patient with this implantable cardioverter defibrillator (icd) heard beeping from the device.Upon analysis of device data, it was found that there was an out-of-range shock lead impedance (sli) measurement of 130 ohms.Technical services (ts) analyzed device data and provided troubleshooting.Reprogramming was performed.The sli returned to within normal limits.No adverse patient effects were reported.Currently, this icd remains in service.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that there was no conclusive evidence that the observed out-of-range impedance measurements were due to a malfunction or an inadequate lead-to-device connection.However, intermittent, out-of-range shock lead impedance measurements with no conclusive evidence of a malfunction or an inadequate lead-to-device connection are likely the result of the low-energy test signal utilized for daily automatic or in-clinic commanded shock lead impedance testing.A software update was released in 2015 to further improve consistency of the impedance test results and provide the clinician with additional diagnostic tools, including programmable shock lead impedance alert limits.
|
|
Event Description
|
It was reported that the patient with this implantable cardioverter defibrillator (icd) heard beeping from the device.Upon analysis of device data, it was found that there was an out-of-range shock lead impedance (sli) measurement of 130 ohms.Technical services (ts) analyzed device data and provided troubleshooting.Reprogramming was performed.The sli returned to within normal limits.No adverse patient effects were reported.Currently, this icd remains in service.
|
|
Search Alerts/Recalls
|
|