• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INCEPTA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INCEPTA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) Back to Search Results
Model Number E161
Device Problems High impedance (1291); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2022
Event Type  malfunction  
Event Description
It was reported that the patient with this implantable cardioverter defibrillator (icd) heard beeping from the device.Upon analysis of device data, it was found that there was an out-of-range shock lead impedance (sli) measurement of 130 ohms.Technical services (ts) analyzed device data and provided troubleshooting.Reprogramming was performed.The sli returned to within normal limits.No adverse patient effects were reported.Currently, this icd remains in service.
 
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that there was no conclusive evidence that the observed out-of-range impedance measurements were due to a malfunction or an inadequate lead-to-device connection.However, intermittent, out-of-range shock lead impedance measurements with no conclusive evidence of a malfunction or an inadequate lead-to-device connection are likely the result of the low-energy test signal utilized for daily automatic or in-clinic commanded shock lead impedance testing.A software update was released in 2015 to further improve consistency of the impedance test results and provide the clinician with additional diagnostic tools, including programmable shock lead impedance alert limits.
 
Event Description
It was reported that the patient with this implantable cardioverter defibrillator (icd) heard beeping from the device.Upon analysis of device data, it was found that there was an out-of-range shock lead impedance (sli) measurement of 130 ohms.Technical services (ts) analyzed device data and provided troubleshooting.Reprogramming was performed.The sli returned to within normal limits.No adverse patient effects were reported.Currently, this icd remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16060667
MDR Text Key308016438
Report Number2124215-2022-55295
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526480768
UDI-Public00802526480768
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960040/S235
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/21/2013
Device Model NumberE161
Device Catalogue NumberE161
Device Lot Number100584
Was Device Available for Evaluation? No
Date Manufacturer Received01/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/13/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexMale
-
-