Model Number 21-7322-24 |
Device Problem
Failure to Deliver (2338)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that a patient presented to the infusion clinic for a pump refill.The pump was on continuous mode, but stopped (with a red bar rather than a green bar) and none of the medication had infused, the iv bag containing the medication was full and the pump indicated that 287.9 ml (of the starting 288 ml) remained to infuse.The battery indicator was full.The patient reports that she did not touch her pump at all and just left it in the fanny pack all night.The nurse notes from the previous day stated that it was verified by two registered nurses as in continuous run.No patient injury was reported.
|
|
Manufacturer Narrative
|
Other, other text: no product was returned.We are unable to confirm the reported complaint.If the product is returned, smiths medical will reopen this complaint for further investigation.
|
|