MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1712K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1712K |
Device Problem
Device Difficult to Program or Calibrate (1496)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/22/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.The insulin pump involved in this event is the ngp 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the ngp insulin pump, which is marketed in the united states.
|
|
Event Description
|
K.N 12/29/2022 information received by medtronic indicated that the customer noticed keypad unresponsive and stated menu button cracked.No harm requiring medical intervention was reported.Troubleshooting was performed, customer will discontinue to use the device.The pump will return for analysis.
|
|
Manufacturer Narrative
|
The pump passed the displacement test and self test.All buttons are functioning properly.The pump was cut open to perform a visual inspection and found no physical or moisture damage to the keypad assembly, electronic assembly, or motor.The j1 connector on pcba 1 was locked properly during visual inspection.A p-cap locks properly into the reservoir compartment.The following were noted during the physical inspection: scratched case, cracked select button keypad overlay, and pillowing keypad overlay.The unresponsive keypad complaint is not confirmed.Cracked select button is confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|