• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS; STOPCOCK, I.V. SET Back to Search Results
Catalog Number 2C6255
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2022
Event Type  malfunction  
Event Description
It was reported that a clearlink system continu-flo solution set leaked at the most proximal valve, closest to the solution.This occurred during priming.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing were performed including pressure test and clear passage and the results were satisfactory.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16068097
MDR Text Key306621983
Report Number1416980-2022-07139
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130245
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C6255
Device Lot NumberR22B02086
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received02/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NA.
-
-