OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", 5 X 330 MM, NEEDLE HOLDER, STRAIGHT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
Model Number WA64700A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2022 |
Event Type
malfunction
|
Event Description
|
The customer contacted the olympus technical assistance center (tac), to report the distal jaw non articulating jaw broke off and fell into the patient during a robotics laparoscopic proctectomy.The customer retrieved the piece that broke off and completed the procedure.No patient injury was reported.
|
|
Manufacturer Narrative
|
In speaking with olympus technical assistance center (tac), the customer declined trouble shooting.The device has not been returned to olympus for evaluation.The investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
|
|
Manufacturer Narrative
|
Updated fields: d8, e4, h4, h6, h10 this report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined as the device was not returned for evaluation.Although the device was not returned, based on the customer's description, it is likely the reported event occurred due wear and tear.Olympus will continue to monitor field performance for this device.
|
|