Catalog Number 410-2000
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2022
Event Type  malfunction  
Event Description
The distributor reported on behalf of the customer that the device, 410-2000 cga, surefit ground pad with 10ft cable, 100/case was being used on approximately (b)(6) 2022 and ¿they have been reported to not stick, leaving skin impressions, and turning brown toward the end of the case.¿ there was no impact or injury to the patient.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Manufacturer Narrative
The device is not expected to be returned for evaluation and review.However, the complaint investigation is not complete at this time.A supplemental and final report will be filed following the completion of the complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Manufacturer Narrative
The device will not be returned, and no photographic evidence was provided.Therefore, the reported event cannot be verified.A two-year lot history review cannot be conducted as a lot number was not provided.A device history record review cannot be conducted as a lot number was not provided.(b)(4).Per the instructions for use, the user is advised the following: prepare the skin at the application site according to facility protocol.If no protocol exists, clip excess hair at application site, clean and disinfect area to remove oils, lotions, etc., and allow to dry thoroughly.Do not open package until ready to apply pad to skin.Inspect the pad and cable.Do not use if product is expired or apparently damaged.Check expiration date on package.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event Description
The distributor reported on behalf of the customer that the device, 410-2000 cga, surefit ground pad with 10ft cabel, 100/case was being used on approximately 25oct2022 and ¿.They have been reported to not stick, leaving skin impressions, and turning brown toward the end of the case.¿.There was no impact or injury to the patient.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CGA, SUREFIT GROUND PAD WITH 10FT CABEL, 100/CASE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX  31136
Manufacturer (Section G)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX   31136
Manufacturer Contact
robin drum
11311 concept blvd
largo, FL 33773
8653881978
MDR Report Key16077558
MDR Text Key307042822
Report Number3007305485-2022-00193
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120322
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number410-2000
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received01/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Patient Sequence Number1
Patient EthnicityNon Hispanic
-
-