MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) JOT DX¿ ICM; IMPLANTABLE CARDIAC MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) JOT DX¿ ICM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM4500
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2022
Event Type  Injury  
Event Description
It was reported the patient presented in clinic for follow-up.During the device check, it was discovered the insertable cardiac monitor (icm) could not be interrogated.No changes or interventions were reported.There were no patient consequences.
 
Event Description
New information noted the insertable cardiac monitor (icm) was explanted.The patient was in stable condition before, during, and after.
 
Manufacturer Narrative
The reported event of failure to interrogate was confirmed.The insertable cardiac monitor (icm) was received for analysis at end of service (eos).A longevity calculation was performed and found the battery depletion was premature.Battery voltage recovery after detachment from device excluded a battery issue for the cause of premature depletion.Current was monitored for an extensive period and no high current drain was noted on hybrid.The cause of premature battery depletion could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOT DX¿ ICM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16080942
MDR Text Key307261376
Report Number2017865-2022-51100
Device Sequence Number1
Product Code MXC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/10/2023
Device Model NumberDM4500
Device Catalogue NumberDM4500
Device Lot NumberP000149188
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
Patient SexMale
-
-