ABBOTT GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDHFA500Q
Device Problems Inappropriate or Unexpected Reset (2959); Interrogation Problem (4017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2022
Event Type  malfunction  
Event Description
The patient presented in-clinic after receiving an auditory alert.Upon investigation, the device was found to be in backup vvi (bvvi) mode and bluetooth low energy (ble) telemetry was noted to be slow.Abbott technical support was contacted and the device was reprogrammed, however, the cause of the bvvi and telemetry anomaly remains unknown.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of bvvi was confirmed.Based on the information provided, it was determined that the bvvi was likely caused by a hardware anomaly.Further investigation could not be performed as the device image could not be obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16087753
MDR Text Key306541025
Report Number2017865-2023-00115
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDHFA500Q
Device Catalogue NumberCDHFA500Q
Device Lot NumberP000146521
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-