Model Number 27977 |
Device Problem
Inappropriate or Unexpected Reset (2959)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
|
Event Description
|
The customer reported that the device was turning on and off.No consequences or impact to patient were reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: other, other text: the returned device was evaluated.The device powered off shortly after being powered on.Shorting inside the on/off power button created an electrical short across the button, resulting in the button being activated even when not physically pressed.
|
|
Event Description
|
The customer reported that the device was turning on and off.No consequences or impact to patient were reported.
|
|