• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION SPECTRUM INFUSION PUMP; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION SPECTRUM INFUSION PUMP; PUMP, INFUSION Back to Search Results
Catalog Number 35700BAX2
Device Problem Intermittent Loss of Power (4016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2022
Event Type  malfunction  
Event Description
During evaluation of a returned spectrum pump, the device was found to power off intermittently.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.Evaluation included a visual assessment as well as functional testing.A service history review revealed no indication that the parts replaced during servicing caused or contributed to this issue.During evaluation, the device was found to power off intermittently.The cause was identified to be a depressed battery pin which requires the rear case to be replaced to correct the condition.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRUM INFUSION PUMP
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEDINA
711 park ave
medina NY 14103
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16124662
MDR Text Key308200816
Report Number1314492-2023-00007
Device Sequence Number1
Product Code FRN
UDI-Device Identifier00085412498683
UDI-Public(01)00085412498683
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133801
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number35700BAX2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
NA.
-
-