• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER/COOLER 3T 110V/60 HZ; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER/COOLER 3T 110V/60 HZ; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2022
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in united states.The livanova field service representative in charge replaced circulation motor due to very loud noise, the patient 2 pump replaced due to not working correctly correcting the ee20 error.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-coolersystem 3t returned error patient circuit 2 pump does not start (e20) during set up.There was no patient involvement.
 
Manufacturer Narrative
Complaints database analysis revealed that no similar event on this device occurred since its installation in 2021.The most likely root cause of the reported issue is a defective patient pump 1 due to motor bearing damaged by the corrosion, which consequently does not allow the motor shaft to rotate.Possible causes are water infiltration or water formation due to condensation or high temperatures and high level of humidity in the stationary phase.Causes related to environmental conditions.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER/COOLER 3T 110V/60 HZ
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key16146702
MDR Text Key308655964
Report Number9611109-2023-00019
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817900528
UDI-Public010403381790052811210326
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-