Model Number 16-02-85 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was not provided.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in united states.The livanova field service representative in charge replaced circulation motor due to very loud noise, the patient 2 pump replaced due to not working correctly correcting the ee20 error.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a heater-coolersystem 3t returned error patient circuit 2 pump does not start (e20) during set up.There was no patient involvement.
|
|
Manufacturer Narrative
|
Complaints database analysis revealed that no similar event on this device occurred since its installation in 2021.The most likely root cause of the reported issue is a defective patient pump 1 due to motor bearing damaged by the corrosion, which consequently does not allow the motor shaft to rotate.Possible causes are water infiltration or water formation due to condensation or high temperatures and high level of humidity in the stationary phase.Causes related to environmental conditions.
|
|
Event Description
|
See initial report.
|
|