Model Number 16-02-80
Device Problem Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(4).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-cooler system 3t tripped the rcd (residual current device) during priming.There was no patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: complaints database analysis revealed that no similar event since unit installation in 2015.The reported event is a known issue.Based on findings, a potential breach of the cooling coil can be generated due to stirrer flow in the tank.Consequently, the water will enter the cooling circuit and the compressor unit.This potential failure mode will cause immediate damage of the cooling compressor system.The compressor failure leads to an increase of the leakage current of the device and the circuit breaker will trip.The erosion kit upgrade was already installed on the device in (b)(6) 2020 and includes the new design of the pump head of the stirrer motor to prevent a water flow directed to a narrow area of the evaporator coil.The issue was claimed about more than year after the upgrade which suggests that the issue is most probably caused by the corrosion and not erosion and occurred overtime independently from the upgrade.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key16155960
MDR Text Key307279208
Report Number9611109-2023-00023
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/28/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/26/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-