LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-80 |
Device Problem
Electrical Shorting (2926)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(4).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t tripped the rcd (residual current device) during priming.There was no patient involvement.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H10: complaints database analysis revealed that no similar event since unit installation in 2015.The reported event is a known issue.Based on findings, a potential breach of the cooling coil can be generated due to stirrer flow in the tank.Consequently, the water will enter the cooling circuit and the compressor unit.This potential failure mode will cause immediate damage of the cooling compressor system.The compressor failure leads to an increase of the leakage current of the device and the circuit breaker will trip.The erosion kit upgrade was already installed on the device in (b)(6) 2020 and includes the new design of the pump head of the stirrer motor to prevent a water flow directed to a narrow area of the evaporator coil.The issue was claimed about more than year after the upgrade which suggests that the issue is most probably caused by the corrosion and not erosion and occurred overtime independently from the upgrade.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Search Alerts/Recalls
|
|
|