• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-712 PARADIGM INSULIN PUMP; PUMP, INFUSION, INSULIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-712 PARADIGM INSULIN PUMP; PUMP, INFUSION, INSULIN Back to Search Results
Model Number MMT-712EWS
Device Problem Date/Time-Related Software Problem (2582)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Medtronic received information that time/clock anomaly occurred.There was no adverse impact or consequence reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MMT-712 PARADIGM INSULIN PUMP
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
7635140379
MDR Report Key16166728
MDR Text Key308727463
Report Number2032227-2023-112966
Device Sequence Number1
Product Code LZG
UDI-Device Identifier000000613994469694
UDI-Public(01)000000613994469694
Combination Product (y/n)N
Reporter Country CodeHK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-712EWS
Device Catalogue NumberMMT-712EWS
Device Lot NumberA8712EWSJ
Was Device Available for Evaluation? No
Date Manufacturer Received03/11/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/18/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-