Model Number 101/860/070
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2022
Event Type  malfunction  
Manufacturer Narrative
Device lot number is unknown, no product information has been provided to date.A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Event Description
It was reported that the pilot balloon was found detached from the inflation line while in use with the patient.No patient injury reported.No additional information available for this complaint.
 
Manufacturer Narrative
One unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.The inflation line was detached from the pilot balloon.What cause the problem is unknown.A dhr (device history review) was not performed because the results of the complaint investigation do not indicate a problem with the manufacture or last repair of the device.No manufacturing or service issues were identified as causes of the customer's reported problem during the review of service and repair records.Therefore no manufacturing or service records review is needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLU SELECT SUCTIONAID TRACHEOSTOMY TUBE
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-1-1 akasaka
minneapolis, MN 55442
MDR Report Key16194904
MDR Text Key308557008
Report Number3012307300-2023-00536
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076134
UDI-Public15019517076134
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/070
Device Catalogue Number101/860/070CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/19/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-