• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDDRA500Q
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hematoma (1884)
Event Date 12/29/2022
Event Type  Injury  
Event Description
Related manufacturing reference number: 2017865-2023-02533.It was reported that the patient presented with a pocket hematoma.Upon interrogation, it was noted that the right ventricular lead exhibited high capture threshold.The physician evacuated the pocket to resolve the hematoma.While the pocket was open, the physician repositioned the right ventricular lead on (b)(6) 2022.While the lead was being repositioned, the helix was difficult to extend.The patient was in stable condition.
 
Event Description
New information received indicates that the right ventricular lead had dislodged.The pacing and defibrillation impedance were low and the lead was not sensing or capturing.Temporary programming changes were made.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16200021
MDR Text Key307657976
Report Number2017865-2023-02535
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberCDDRA500Q
Device Catalogue NumberCDDRA500Q
Device Lot NumberS000088640
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/18/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2088TC/46 TENDRIL STS LEAD
Patient Outcome(s) Other; Required Intervention;
Patient Age62 YR
Patient SexFemale
-
-