COMPANION MEDICAL INC INPEN MMT-105NNGYNA NOVO NORDISK GRAY; SYRINGE, PISTON Back to Search Results
Model Number MMT-105NNGYNA
Device Problems Break (1069); Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2022
Event Type  malfunction  
Manufacturer Narrative
Per visual inspection: cap does not fit securely to inpen.No physical damage to cartridge holder was noted.Unit paired successfully to commercial app.Inpen received with leadscrew 1/4 of the travel.Re-wound screw.No drag was observed, the screw was not bent, advanced when dosage knob was pressed dialing a dosage and retracted appropriately.No resistance was observed when dosing without a cartridge installed.The screw advanced every time 30.0u was dialed and dosed until the screw reached max extension.Installed front shell to testing inpen and front shell did not stay attached.Inpen front shell does not fit securely onto cartridge holder due to small snap arm being cracked / broken.In conclusion: inpen cap does not fit securely onto cartridge holder due to small snap arm being cracked / broken.The customer complaint of cap no longer staying attached was confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information received by medtronic indicated that the insulin pens were broken and damaged and was not working.Customer stated that cap was no longer attached.No harm requiring medical intervention was reported.The device was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNGYNA NOVO NORDISK GRAY
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
tricha miles
1223 world trade drive, 100
san diego, CA 92128
7635140379
MDR Report Key16221761
MDR Text Key308371392
Report Number3012822846-2023-00494
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000351
UDI-Public(01)000010862088000351(17)230422
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/22/2023
Device Model NumberMMT-105NNGYNA
Device Catalogue NumberMMT-105NNGYNA
Device Lot NumberB0082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2023
Date Manufacturer Received12/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age58 YR
Patient SexMale
Patient Weight86 KG
-
-